Model L20
Instruction Manual
Microprocessor based
pH, ORP, Temperature
Bench Top Meter
Electro-Chemical Devices, Inc.
1681 Kettering, Irvine, CA 92614 USA
Telephone: +1-949-336-6060
FAX: +1-949-336-6064
www.ecdi.com
Seitenansicht 0
1 2 3 4 5 6 ... 19 20

Inhaltsverzeichnis

Seite 1 - Instruction Manual

Model L20 Instruction Manual Microprocessor based pH, ORP, Temperature Bench Top Meter Electro-Chemical Devices, Inc. 1681 Kettering, Irvine, CA 92

Seite 2

4. Rinse the pH and ATC/Temp probe in distilled water and immerse them in the second buffer solution (either 4.00/4.01 or 9.18/10.01). Allow temperat

Seite 3 - Initial Inspection

then press “Slope” key to calibrate. Once the unit calibrates the second point, the “SLOPE” icon will flash. The unit is ready to be sloped at the thi

Seite 4

track the pH value as sensed by the pH electrode and the ATC/Temp probe. 5. When the “WAIT” icon disappears, the reading is then “LOCK” and will not

Seite 5 - A. Display

3. Rinse the pH electrode and ATC/temp probe with distilled water and immerse in the sample to be measured. 4. Allow sufficient time for the display

Seite 6

[Note: For samples that are inherently unstable, the unit will not AUTOLOCK. In this case, use the pH NON- AUTOLOCK mode for measurements.] Measuremen

Seite 7

Error Displays and troubleshooting Main Display Possible Causes Corrective Actions “Er1” ● “Stand” was pressed before the electrode and ATC/Temp probe

Seite 8

Specifications Display Range Resolution Accuracy pH -2.00 to 16.00 pH 0.01 pH ±0.1% FS mV -1999 to 1999 mV 1 mV ±0.1% FS Temperature 0.0 to 100.0 °C 0

Seite 10

Warranty Electro-Chemical Devices, Inc. (ECD) warrants all products it manufactures to be free from defect in materials and factory workmanship, and a

Seite 12

Contents General Introduction………………………. 3 Initial Inspection……………………………... 3 Powering the Unit……………………………. 4 Installing the Batteries…………………….. 4 Dis

Seite 13

1681 Kettering, Irvine, CA USA Phone +1 949 336-6060 FAX +1 949 336-6064 www.ecdi.com 20

Seite 14

General Introduction Thank you for selecting the Model L20 pH meter. The Model L20 is a precision tool that measures pH, mV, ORP and temperature. A b

Seite 15 - correct measuring

Powering the Unit The model pH6175 can be powered by an 115V or 230VAC adaptor as well as 6 “AAA” alkaline batteries. Check the label on the AC 3 adap

Seite 16 - Specifications

Display and Key Functions A. Display Figure 2 1. WAIT This will be displayed when the unit is still waiting for a stable reading or end point sensin

Seite 17

B. Keys On/Off Press and hold this key for 5 seconds to power on and shut off the meter. Once the unit is power up, press the same key to turn on or

Seite 18 - Warranty

Operational Procedures A. Buffer Set Selection The Model L20 meter has two buffer sets: 7.00, 4.01, 10.01pH and 6.86, 4.00, 9.18 pH. The meter is fact

Seite 19

4. Rinse the pH and ATC/Temp probe in distilled water and immerse them in the second buffer solution (either 4.00/4.01 or 9.18/10.01). Allow temperat

Seite 20

buffers start to flash. The unit is ready to be sloped at the second buffer. 4. Rinse the pH probe in distilled water and immerse it in the second bu